Basic medical laboratory, ethics and microscopes

18 Kerala MSc MLT Entrance practice questions on Basic medical laboratory, ethics and microscopes, part of Pathology. The first 18 are below; the first 5 with answers and explanations (tap Show answer after you have chosen), the rest in Learn with a free account.

  1. Question 1 · Medium

    What does setting up Köhler illumination on a bright-field microscope mainly give?

    1. AA dark background against which unstained cells shine
    2. BA higher total magnification with the same lenses
    3. CAn evenly lit field with the best resolution and contrast
    4. DFluorescent light from the specimen itself
    Show answer

    Answer: C · An evenly lit field with the best resolution and contrast

    In Köhler illumination the field diaphragm is focused in the specimen plane by the condenser, and the condenser is centred. The lamp filament is then out of focus in the specimen plane, so the field is evenly lit without glare. The aperture diaphragm is then opened to about 70-80% of the objective's numerical aperture, which gives the best balance of resolution and contrast. Magnification is not changed.

    • A: A dark background is the purpose of dark-field illumination.
    • B: Magnification depends only on the objective and eyepiece.
    • D: Fluorescence needs an excitation source and filters.
  2. Question 2 · Medium

    Blood is drawn from a driver for a medico-legal blood alcohol estimation. What should the venepuncture site be cleaned with?

    1. AAqueous povidone-iodine or soap and water
    2. BRectified spirit on a cotton swab
    3. CA 70% isopropyl alcohol swab
    4. DEther, allowed to evaporate fully
    Show answer

    Answer: A · Aqueous povidone-iodine or soap and water

    Any alcohol-based skin antiseptic can carry alcohol into the sample and falsely raise the result, which a defence lawyer can challenge. A non-alcoholic antiseptic such as aqueous povidone-iodine, chlorhexidine in water or soap and water is used. The sample is collected in a fluoride tube, sealed, labelled and handed over with a signed chain of custody. These steps make the result legally defensible.

    • B: Rectified spirit is ethanol, the very substance being measured.
    • C: Isopropyl alcohol can contaminate the sample and interfere with the result.
    • D: Ether is volatile and flammable and is not a recommended skin preparation for this test.
  3. Question 3 · Easy

    Explaining the purpose, benefits and risks of an HIV test to a patient and getting the patient's agreement before the sample is taken is called what?

    1. AConfidentiality
    2. BBeneficence
    3. CPre-test screening
    4. DInformed consent
    Show answer

    Answer: D · Informed consent

    Informed consent means the patient is told what the test is for, its benefits and risks and the alternatives, and agrees freely before it is done. Indian HIV testing policy (NACO) requires consent, counselling and confidentiality for every HIV test. Consent is recorded, and for minors it is given by a guardian. Testing without it is an ethical and legal breach.

    • A: Confidentiality is keeping the result private, not obtaining agreement.
    • B: Beneficence is the duty to act for the patient's good.
    • C: Screening is a type of testing, not the process of agreement.
  4. Question 4 · Hard

    Useful magnification is about 500-1,000 times the numerical aperture. With a 40x objective of NA 0.65, which eyepiece gives the highest magnification without empty magnification?

    1. A25x
    2. B20x
    3. C15x
    4. D10x
    Show answer

    Answer: C · 15x

    The upper limit is 1,000 x 0.65 = 650x. With a 15x eyepiece, the total is 40 x 15 = 600x, which is within the limit. A 20x eyepiece gives 800x and a 25x gives 1,000x: the image grows but no finer detail is resolved (empty magnification). A 10x eyepiece gives 400x, which is useful but not the highest possible.

    • A: A 25x eyepiece gives 1,000x, well into empty magnification.
    • B: A 20x eyepiece gives 800x, above the 650x limit.
    • D: A 10x eyepiece gives 400x, below the highest useful value.
  5. Question 5 · Hard

    After examining oil-immersion slides, a technician turns to the 40x dry objective. At 40x the image is milky and will not sharpen, while the 10x and 100x images of the same slide are clear. The most likely cause is:

    1. ASlide placed on the stage upside down
    2. BLamp voltage set too low for 40x
    3. CImmersion oil smeared on the 40x front lens
    4. DCondenser iris diaphragm closed too far
    Show answer

    Answer: C · Immersion oil smeared on the 40x front lens

    The problem is confined to one objective, so it lies in that objective. The 40x lens has a short working distance and often dips into oil when the nosepiece is swung through an oiled slide. A film of oil on its front lens gives a milky, unsharp image. Cleaning it with lens tissue and lens cleaner restores the image.

    • A: An upside-down slide also stops the 100x objective from focusing.
    • B: A dim lamp would affect all objectives and darken rather than blur the image.
    • D: A closed iris would affect every objective and gives a dark, contrasty image rather than milkiness.
  6. Question 6 · Medium

    A patient's brother phones the laboratory and asks for the patient's HIV test result. What should the technologist do?

    1. AGive it only after confirming his identity and relationship
    2. BGive it after confirming the patient's age
    3. CDecline, and release it only to the patient or the doctor
    4. DGive it only if the result is negative

    Answer and explanation in LearnFree account

  7. Question 7 · Medium

    In a humid coastal laboratory, how should microscopes be stored to prevent fungal growth on the lenses?

    1. AIn the laboratory cold room
    2. BIn a dry cabinet with silica gel or a warm bulb
    3. CSealed airtight in a plastic bag
    4. DWith the eyepieces removed to air the tubes

    Answer and explanation in LearnFree account

  8. Question 8 · Easy

    Objectives that stay nearly in focus when the nosepiece is turned from one to the next are described as:

    1. AAchromatic
    2. BParfocal
    3. CApochromatic
    4. DParcentric

    Answer and explanation in LearnFree account

  9. Question 9 · Easy

    When focusing with the 100x oil-immersion objective, only the fine adjustment is used because:

    1. AThe condenser must stay fixed
    2. BThe coarse knob does not move the stage
    3. CIts working distance is very short
    4. DOil blocks the coarse adjustment

    Answer and explanation in LearnFree account

  10. Question 10 · Hard

    A technician pricks a finger with a needle just used to draw blood from a patient known to be HIV-positive. Which plan follows NACO guidance?

    1. AStart PEP only after the technician's own baseline HIV result is back
    2. BSqueeze the wound to bleed it, apply bleach and start PEP after 72 hours
    3. CWait for the patient's repeat HIV test before washing the site
    4. DWash with soap and water without squeezing; start PEP as early as possible

    Answer and explanation in LearnFree account

  11. Question 11 · Medium

    Studies of laboratory errors consistently show that most of them arise in which phase of testing?

    1. APre-analytical phase
    2. BPost-analytical phase
    3. CAnalytical phase
    4. DInstrument calibration phase

    Answer and explanation in LearnFree account

  12. Question 12 · Medium

    Hyaline casts and unstained living cells are easier to see with a phase-contrast microscope because it:

    1. ALights the specimen only from the side on a dark field
    2. BIncreases magnification beyond that of bright-field lenses
    3. CConverts phase differences in light into differences in brightness
    4. DUses ultraviolet light that makes the specimen glow

    Answer and explanation in LearnFree account

  13. Question 13 · Medium

    A centrifuge motor catches fire in the laboratory. After switching off the power at the mains, which extinguisher is the safest to use?

    1. ACarbon dioxide
    2. BWater
    3. CSoda-acid
    4. DMechanical foam

    Answer and explanation in LearnFree account

  14. Question 14 · Hard

    A technician sees a dark speck in the field at every magnification. It does not move when the slide is moved, and it turns round when the eyepiece is rotated in its tube. Where is the dirt?

    1. AOn the condenser top lens
    2. BOn the eyepiece lens
    3. COn the slide or coverslip
    4. DOn the front lens of the objective

    Answer and explanation in LearnFree account

  15. Question 15 · Easy

    The limit of resolution of a good bright-field light microscope using oil immersion is about:

    1. A0.02 µm
    2. B0.2 µm
    3. C20 µm
    4. D2 µm

    Answer and explanation in LearnFree account

  16. Question 16 · Medium

    A conscious adult patient, told the purpose and risks of a blood test, refuses it. The laboratory respects the refusal. This follows the ethical principle of:

    1. ABeneficence
    2. BJustice
    3. CAutonomy
    4. DNon-maleficence

    Answer and explanation in LearnFree account

  17. Question 17 · Medium

    For a medico-legal specimen, the documented record of every person who collected, handled, stored and tested it is called the:

    1. AInformed consent record
    2. BInternal quality control log
    3. CChain of custody
    4. DRequisition form

    Answer and explanation in LearnFree account

  18. Question 18 · Hard

    A stained blood film is in sharp focus under the 10x objective, but with the 100x oil objective no image can be found even when the lens touches the oil. The oil and objective are clean. What is the most likely cause?

    1. AToo much immersion oil has been applied
    2. BThe slide has been placed upside down on the stage
    3. CThe 100x objective is not parfocal with the others
    4. DThe condenser iris diaphragm is fully open

    Answer and explanation in LearnFree account