Question 1 · Medium
A solvent sold as spectroscopic (or HPLC) grade is chosen for photometric work mainly because it has:
- AVery low absorbance in the UV range
- BA higher boiling point than usual
- CA lower price than analytical grade
- DA built-in pH buffer
Show answerHide answer
Answer: A · Very low absorbance in the UV range
Spectroscopic and HPLC grade solvents are purified to remove traces that absorb ultraviolet light, and the label states the wavelength below which the solvent itself absorbs (its UV cut-off). A solvent with absorbing impurities raises the blank and the baseline, so readings at low wavelengths become unreliable. These grades cost more than ordinary analytical grade. Their boiling point and pH are those of the pure solvent.
- B: Boiling point is a property of the solvent, not of its purity grade.
- C: Spectroscopic grade is more expensive, not cheaper, than analytical grade.
- D: These solvents contain no added buffer.