Blood bank reagents and quality control

9 DOH Abu Dhabi (HAAD) lab licensing practice questions on Blood bank reagents and quality control, part of Blood Banking. The first 9 are below; the first 5 with answers and explanations (tap Show answer after you have chosen), the rest in Learn with a free account.

  1. Question 1 · Easy

    Unless the package insert states otherwise, antisera and reagent red cells are stored at:

    1. A-20 °C or colder
    2. B35 to 37 °C
    3. C20 to 24 °C
    4. D2 to 8 °C
    Show answer

    Answer: D · 2 to 8 °C

    Blood bank reagents are refrigerated at 2 to 8 °C and must not be frozen, which would lyse reagent red cells. Keeping them at room temperature speeds the loss of antibody activity and antigen strength and lets bacteria grow. Reagents are taken out only for use and returned promptly. Refrigerator temperatures are monitored and recorded as part of quality control.

    • A: Freezing lyses reagent red cells and is not used for routine reagents.
    • B: 37 °C is the incubation temperature for tests, not a storage temperature.
    • C: Room temperature speeds deterioration and bacterial growth.
  2. Question 2 · Medium

    Reagent red cells have a much shorter shelf life than most antisera. The main reason is that during storage:

    1. ARed cell antigens weaken and the cells slowly hemolyze
    2. BThe cells become coated with IgG from the preservative
    3. CRed cell antigens grow stronger and cause false positives
    4. DThe suspension concentrates as the water evaporates
    Show answer

    Answer: A · Red cell antigens weaken and the cells slowly hemolyze

    Reagent red cells are cells kept in a preservative solution, and some of their antigens weaken during storage. Aging cells also hemolyze, and weakened antigens can lead to false-negative antibody screens. Antisera are proteins that stay stable much longer when refrigerated. Daily QC and using cells within their dating guard against these losses.

    • B: The preservative contains no IgG and does not coat the cells.
    • C: Antigens weaken, not strengthen, as reagent cells age.
    • D: Evaporation is not the reason for the short dating.
  3. Question 3 · Easy

    For tube agglutination tests, red cells are usually suspended in saline at a concentration of:

    1. A2 to 5%
    2. B40 to 50%
    3. C20 to 25%
    4. D0.8 to 1%
    Show answer

    Answer: A · 2 to 5%

    A 2 to 5% suspension gives the right balance of antigen and antibody for tube tests. Too many cells can mask weak reactions, and too few make the cell button hard to read. Column agglutination (gel) cards use a weaker suspension of about 0.8 to 1%. Commercial reagent red cells are supplied ready to use at the strength stated in the package insert.

    • B: A 40 to 50% suspension is close to whole blood and would mask weak reactions.
    • C: A 20 to 25% suspension has far too many cells for tube tests.
    • D: 0.8 to 1% is the suspension used in column agglutination (gel) tests.
  4. Question 4 · Medium

    A technologist interprets today's panel using the antigen sheet (antigram) from the previous lot of panel cells. What is the main risk?

    1. AThe reaction strengths will be graded too high
    2. BNo risk, as all lots carry the same donor phenotypes
    3. CWrong rule-outs, because cell phenotypes differ between lots
    4. DOnly the autocontrol result becomes invalid
    Show answer

    Answer: C · Wrong rule-outs, because cell phenotypes differ between lots

    Each lot of panel cells comes from different donors, so cell 3 in one lot has no relation to cell 3 in the next. Reading reactions against the wrong antigram can exclude an antibody that is present or name one that is not. The antigram's lot number must match the vials in use and should be checked before interpretation. Recording the lot number on the worksheet makes this check routine.

    • A: Grading of reactions does not depend on the antigen sheet.
    • B: Panel lots come from different donors, so their phenotypes differ.
    • D: The autocontrol uses the patient's own cells and does not depend on the antigram.
  5. Question 5 · Medium

    A vial of antibody screening cells, within its expiry date, has a dark red supernatant, and the cells look clumped after gentle mixing. What should be done?

    1. AUse it only for patients with no antibodies
    2. BDiscard it and use a fresh vial
    3. CUse it, as the antigens are unaffected
    4. DWash the cells, then use them
    Show answer

    Answer: B · Discard it and use a fresh vial

    Marked hemolysis or clumping that does not disperse suggests bacterial contamination or deterioration, even before the expiry date. Damaged cells can lose antigen strength and give false-negative screens, or clump and give false positives. Every reagent is inspected before use, and an unacceptable one is discarded and the problem recorded. Results obtained with it should be reviewed.

    • A: A false-negative screen would harm exactly the patients whose antibodies are missed.
    • C: Hemolysis and clumping can come with loss of antigen strength.
    • D: Washing does not restore damaged antigens or remove contamination.
  6. Question 6 · Medium

    For daily QC of anti-Jka antiserum, which positive control cell gives the best check of reagent strength?

    1. AA Jk(a+b-) cell, with a double dose of Jka
    2. BA Jk(a-b-) cell, which lacks both antigens
    3. CA Jk(a+b+) cell, with a single dose of Jka
    4. DA Jk(a-b+) cell, which lacks Jka

    Answer and explanation in LearnFree account

  7. Question 7 · Hard

    Daily QC of the tube LISS antibody screen uses a weak anti-D control serum. Today the D-positive screening cells give 0 with it at the AHG phase, and a new vial of control serum gives the same result. Check cells added to these tubes react 2+. The most likely cause is:

    1. AA fault in the sensitization (37 °C incubation) step
    2. BAn inactive anti-IgG reagent
    3. CContamination of AHG with human serum
    4. DPoor washing that neutralized the AHG

    Answer and explanation in LearnFree account

  8. Question 8 · Easy

    IgG-sensitized check cells used in antiglobulin testing are usually made by coating:

    1. AGroup O red cells with IgM anti-A
    2. BGroup A red cells with complement only
    3. CD-positive red cells with IgG anti-D
    4. DD-negative red cells with IgG anti-D

    Answer and explanation in LearnFree account

  9. Question 9 · Easy

    Daily quality control of an anti-D typing reagent should include:

    1. AThe first patient sample of the day
    2. BA D-positive cell and a D-negative cell
    3. CTwo D-positive cells of different strength
    4. DA D-negative cell only

    Answer and explanation in LearnFree account

More Blood Banking questions

Answer them one by one, keep your score and see your weak topics in Learn. Signed-out visitors try 25 per exam; sign in with Google to see them all.

Study Blood Banking