Question 1 · Hard
A 41-year-old man develops fever and a widespread rash 4 weeks after starting carbamazepine. The analyzer's differential is suppressed by an abnormal-scatter flag, so a direct eosinophil count is requested. Blood is diluted 1:20 in an eosin-based fluid that lyses the red cells and stains eosinophil granules. In an improved Neubauer chamber (depth 0.1 mm), all nine large squares (1 mm² each) are counted on both sides, giving 41 and 49 eosinophils. The reference value is below 0.5 × 10^9/L. What is the eosinophil count, and how is it interpreted?
- A1.0 × 10^9/L, an eosinophilia
- B0.10 × 10^9/L, within the reference range
- C2.0 × 10^9/L, an eosinophilia
- D0.05 × 10^9/L, within the reference range
Show answerHide answer
Answer: A · 1.0 × 10^9/L, an eosinophilia
The volume counted is 18 mm² × 0.1 mm = 1.8 µL, so the count is 90 × 20 ÷ 1.8 = 1,000 cells/µL, or 1.0 × 10^9/L. This is twice the upper reference value, an eosinophilia consistent with a drug hypersensitivity reaction. Dividing the cells from both sides by the volume of one side (0.9 µL) doubles the answer, leaving out the 0.1 mm depth gives a result ten times too low, and leaving out the dilution gives 0.05 × 10^9/L. A direct chamber count avoids multiplying a flagged, imprecise eosinophil percentage by the WBC.
- B: This leaves out the 0.1 mm chamber depth.
- C: This divides both sides' cells by one side's volume (0.9 µL).
- D: This leaves out the 1:20 dilution factor.