  1. Question 1 · Hard

    Glucose by GOD-POD is read against distilled water: reagent blank 0.05, standard (100 mg/dL) 0.45 and test 0.65. What is the glucose concentration?

    1. A144 mg/dL
    2. B67 mg/dL
    3. C133 mg/dL
    4. D150 mg/dL
    Show answer

    Answer: D · 150 mg/dL

    Subtract the reagent blank from both readings: test 0.65 - 0.05 = 0.60 and standard 0.45 - 0.05 = 0.40. Glucose = (0.60 / 0.40) x 100 = 150 mg/dL. Leaving the blank in gives 144 mg/dL, and inverting the ratio gives 67 mg/dL. Setting the instrument to zero with the reagent blank instead of water removes this step.

    • A: 144 mg/dL comes from not subtracting the blank (0.65/0.45 x 100).
    • B: 67 mg/dL comes from inverting the ratio (0.40/0.60 x 100).
    • C: 133 mg/dL comes from correcting only the test reading (0.60/0.45 x 100).
  2. Question 2 · Medium

    In a fluorimeter, the emitted light is measured at a right angle to the exciting beam. Why?

    1. ATo increase the wavelength of the emitted light
    2. BTo remove the need for an emission filter
    3. CTo allow two samples to be read together
    4. DTo keep the exciting light out of the detector
    Show answer

    Answer: D · To keep the exciting light out of the detector

    Fluorescent molecules absorb light of one wavelength and emit light of a longer wavelength in all directions. With the detector at 90 degrees, the strong transmitted exciting beam does not reach it, so mainly emitted light is measured. An emission filter or monochromator removes any scattered exciting light. This arrangement gives fluorimetry much greater sensitivity than absorption photometry.

    • A: The emission wavelength is set by the molecule, not by the detector angle.
    • B: A fluorimeter still needs both an excitation and an emission filter or monochromator.
    • C: The right-angle geometry serves one sample beam; it does not read two samples.
  3. Question 3 · Easy

    In a flame photometer, sodium emits light of which color?

    1. ABrick red
    2. BYellow
    3. CCrimson red
    4. DViolet
    Show answer

    Answer: B · Yellow

    Excited sodium atoms emit light at 589 nm, which is yellow, and the instrument reads this line through a matching filter. Potassium gives a lilac (violet) flame and is read at 766 nm, lithium gives crimson red and calcium brick red. Flame photometers are used mainly for sodium and potassium in serum and urine, and many use lithium as an internal standard.

    • A: Brick red is the flame color of calcium.
    • C: Crimson red is the flame color of lithium.
    • D: Violet (lilac) is the flame color of potassium.
  4. Question 4 · Easy

    In atomic absorption spectrophotometry, what is the cathode of the hollow cathode lamp made of?

    1. ADeuterium for all elements
    2. BMercury for all elements
    3. CTungsten for all elements
    4. DThe element being measured
    Show answer

    Answer: D · The element being measured

    The lamp's cathode is made of, or coated with, the same metal that is to be measured, so it emits that element's exact resonance lines. Ground-state atoms of the element in the flame absorb these lines, and the absorbance is proportional to concentration. This gives AAS high specificity, but a separate lamp is needed for each element, such as calcium, magnesium, lead, zinc or copper.

    • A: A deuterium lamp gives continuous ultraviolet light for UV spectrophotometry.
    • B: A mercury vapor lamp gives mercury lines only and is used as a source in some fluorimeters.
    • C: A tungsten lamp gives continuous visible light for colorimeters and spectrophotometers.
  5. Question 5 · Medium

    A colorimetric test gives a red solution. Which filter color should be selected to read it?

    1. AGreen
    2. BYellow
    3. CRed
    4. DOrange
    Show answer

    Answer: A · Green

    A solution looks red because it transmits red light and absorbs its complementary color, green. The filter must pass the light that the solution absorbs most, so a green filter gives the greatest change in absorbance with concentration. Red, orange or yellow light passes through the solution, so readings change little and sensitivity is poor. In practice, the best filter is the one that gives the highest absorbance against the blank.

    • B: Yellow light is also poorly absorbed by a red solution.
    • C: A red solution transmits red light, so a red filter gives almost no absorbance.
    • D: Orange light is largely transmitted by a red solution, giving poor sensitivity.
  6. Question 6 · Medium

    Atomic absorption spectrophotometry differs from flame emission photometry in that it measures light:

    1. AScattered by suspended particles
    2. BEmitted by fluorescent molecules
    3. CAbsorbed by ground-state atoms
    4. DEmitted by excited atoms

    Answer and explanation in LearnFree account

  7. Question 7 · Easy

    A centrifuge rotor whose tube holders lie level during the spin, so the sediment packs flat at the tube bottom, is called a:

    1. AVertical rotor
    2. BZonal rotor
    3. CFixed-angle rotor
    4. DSwing-out rotor

    Answer and explanation in LearnFree account

  8. Question 8 · Easy

    Firefly luciferase reacts with ATP and gives off light. Which instrument measures this light?

    1. AFluorimeter
    2. BLuminometer
    3. CNephelometer
    4. DFlame photometer

    Answer and explanation in LearnFree account

  9. Question 9 · Medium

    A spectrophotometer is more specific than a filter colorimeter mainly because its prism or grating:

    1. ASelects a narrow band of wavelengths
    2. BIncreases the intensity of the light
    3. CLengthens the light path in the cuvette
    4. DRemoves the need for a reagent blank

    Answer and explanation in LearnFree account

  10. Question 10 · Hard

    A pH meter is calibrated with pH 4.00 and 7.00 buffers and then used to adjust a carbonate-bicarbonate buffer to pH 10.0. The buffer later performs poorly. What should be done before the next batch?

    1. ACalibrate with the pH 7.00 buffer alone
    2. BRecalibrate with fresh 4.00 and 7.00 buffers
    3. CCalibrate with pH 7.00 and 10.00 buffers
    4. DCheck the pH with universal indicator paper

    Answer and explanation in LearnFree account

  11. Question 11 · Easy

    A magnetic stirrer mixes a solution by:

    1. ASpinning a coated magnetic bar inside the vessel
    2. BShaking the tube in fast small circles
    3. CTilting the flask slowly from side to side
    4. DPassing ultrasound waves through the liquid

    Answer and explanation in LearnFree account

  12. Question 12 · Easy

    In a colorimeter, the absorbance of a coloured solution in a fixed cuvette rises in a straight line with its concentration. This relationship is the:

    1. AGraham's law
    2. BBeer-Lambert law
    3. CBoyle's law
    4. DStokes law

    Answer and explanation in LearnFree account

  13. Question 13 · Medium

    In fluorimetry, compared with the exciting light, the light emitted by the sample has:

    1. AThe same wavelength
    2. BA shorter wavelength
    3. CNo fixed wavelength
    4. DA longer wavelength

    Answer and explanation in LearnFree account

  14. Question 14 · Hard

    A standard of 10 mg/dL reads 50% transmittance and a test reads 25% transmittance in the same colorimeter. What is the test concentration? (log 2 = 0.30)

    1. A20 mg/dL
    2. B40 mg/dL
    3. C5 mg/dL
    4. D15 mg/dL

    Answer and explanation in LearnFree account

  15. Question 15 · Easy

    A nephelometer measures:

    1. ALight given off by a chemical reaction
    2. BLight scattered by particles in the solution
    3. CLight absorbed by a coloured solution
    4. DLight emitted by atoms in a flame

    Answer and explanation in LearnFree account

  16. Question 16 · Medium

    Many flame photometers add lithium to every standard and sample as an internal standard. What does it correct for?

    1. AHemolysis of the serum sample
    2. BErrors in the lamp output
    3. CChanges in flame and aspiration rate
    4. DProtein in the serum sample

    Answer and explanation in LearnFree account

  17. Question 17 · Easy

    In a laboratory pH meter, which electrode responds to the hydrogen ion concentration of the sample?

    1. ASilver-silver chloride electrode
    2. BPlatinum electrode
    3. CCalomel electrode
    4. DGlass electrode

    Answer and explanation in LearnFree account

  18. Question 18 · Medium

    A test tube reads absorbance 1.9, far above the top standard, on a method that is linear only to 1.0. What should be done?

    1. ADilute the sample and repeat the reading
    2. BReport the result from the calculation
    3. CRead the tube again against water
    4. DChange to a filter of another colour

    Answer and explanation in LearnFree account

  19. Question 19 · Hard

    During a run on a flame photometer, the readings of the standards, controls and patient sera all fall steadily, and the flame flickers unevenly when the sample is sucked in. The most likely cause is:

    1. AExpired control material
    2. BA partly blocked aspirator capillary
    3. CWrong filter for the element
    4. DHemolysis in the patient samples

    Answer and explanation in LearnFree account

  20. Question 20 · Hard

    A serum sample is diluted 1 in 5 because its colour was too deep. The diluted sample reads absorbance 0.40 and a 50 mg/dL standard reads 0.50, both against a reagent blank. What is the result to report?

    1. A312 mg/dL
    2. B40 mg/dL
    3. C200 mg/dL
    4. D8 mg/dL

    Answer and explanation in LearnFree account

  21. Question 21 · Medium

    Serum IgG, IgA, IgM and CRP are measured in many laboratories by which instrument?

    1. AFluorimeter
    2. BAtomic absorption spectrophotometer
    3. CFlame photometer
    4. DNephelometer

    Answer and explanation in LearnFree account

  22. Question 22 · Easy

    Which lamp supplies ultraviolet light in a UV-visible spectrophotometer?

    1. ASodium vapour lamp
    2. BTungsten lamp
    3. CDeuterium lamp
    4. DHollow cathode lamp

    Answer and explanation in LearnFree account

  23. Question 23 · Medium

    In atomic absorption measurement of serum calcium, lanthanum chloride is added to the diluent in order to:

    1. AAct as the light source
    2. BRaise the flame temperature
    3. CRemove protein from the sample
    4. DRelease calcium from phosphate

    Answer and explanation in LearnFree account

  24. Question 24 · Medium

    Why must quartz or silica cuvettes be used for readings at 340 nm?

    1. AQuartz removes stray visible light
    2. BOrdinary glass absorbs ultraviolet light
    3. CQuartz keeps the solution at 37 °C
    4. DGlass reacts with NADH in the sample

    Answer and explanation in LearnFree account