Question 1 · Medium
A type and screen tube shows the patient's first name spelled differently from the request; the medical record number matches. What should the laboratory do?
- APhone the ward and accept on verbal confirmation
- BReject it and collect a new sample
- CCorrect the spelling on the tube label
- DAccept it, since the record number matches
Show answerHide answer
Answer: B · Reject it and collect a new sample
Any discrepancy between the tube and the request means the sample's identity cannot be trusted. The sample is rejected and a new one is collected with correct identification at the bedside. Relabeling a pretransfusion sample is not allowed. Strict rules prevent wrong-blood-in-tube errors, a leading cause of ABO-incompatible transfusion.
- A: Verbal confirmation cannot fix a labeling discrepancy.
- C: Relabeling a pretransfusion sample is not allowed.
- D: One matching identifier does not make up for a discrepancy.