Question 1 · Hard
A 38-year-old man's vasectomy specimen contains two pieces of tubular tissue, each about 1 cm long. At grossing, the pathologist cuts a thin transverse disc from each end of each piece, places the four discs in one cassette and writes 'embed cut faces down' on the cassette request. The H&E slide shows four long strips of smooth muscle and connective tissue, with no lumen and no lining epithelium in any of them, and two deeper levels look the same. The pathologist needs complete cross sections showing the whole ring of each piece. What should be done?
- AMelt out the discs and re-embed them cut face down
- BRotate the block 90° in the clamp and recut it
- CRecut the same block at 2 µm instead of 4 µm
- DCut more deeper levels from the existing block
Show answerHide answer
Answer: A · Melt out the discs and re-embed them cut face down
A transverse disc has two flat cut faces that each show the whole ring of the tube, so sections cut parallel to those faces are cross sections with the lumen in the middle. Long strips of muscle with no lumen mean the discs were laid on their curved sides, so the knife is cutting along the length of the wall. The only fix is to melt the discs out of the wax and re-embed them flat, cut face down, pressed evenly to the bottom of the mold. Deeper levels stay in the same wrong plane, turning the block in the clamp changes only the direction the knife travels across the same face, and a thinner section does not change the plane at all. Orientation notes on the cassette must be read before each piece is placed in the mold.
- B: Turning the block in the clamp changes the cutting direction, not the plane of the section.
- C: Section thickness does not change the plane in which the tissue is cut.
- D: Deeper levels stay in the same longitudinal plane and cannot give cross sections.