Question 1 · Hard
An automated stainer runs H&E on 400 to 500 slides a day. This week, many sections show irregular pale patches that have taken up neither hematoxylin nor eosin, while the rest of the tissue on the same slide stains normally, with blue nuclei and pink cytoplasm. The patches fall in different places on serial sections from the same block. The stainer log shows: hematoxylin replaced this morning; acid alcohol and bluing reagent changed daily; the two dewaxing xylene stations at the start of the program last changed 6 days ago, with a limit of 1,000 slides per change. Slide drying before staining is unchanged. The most likely cause is:
- ABluing step too short after acid differentiation
- BWater carried over into the final clearing xylene
- CHematoxylin exhausted by the high slide throughput
- DWax left in the sections by overused dewaxing xylene
Show answerHide answer
Answer: D · Wax left in the sections by overused dewaxing xylene
Patches that take up neither dye point to a problem before staining starts, because hematoxylin and eosin are applied in aqueous or alcoholic solutions that cannot enter tissue still holding wax. At about 450 slides a day, the dewaxing xylene has handled roughly 2,700 slides in 6 days, nearly three times its limit, so it is loaded with paraffin and leaves wax behind in random areas. Changing the dewaxing stations and keeping to the slide limit fixes the problem, and affected cases are recut or restained. Worn-out hematoxylin gives pale nuclei with normal eosin, short bluing leaves nuclei reddish over the whole section, and water in the final xylene gives a milky haze after coverslipping rather than unstained patches.
- A: Short bluing leaves nuclei reddish across the whole section; it does not leave unstained patches.
- B: Water in the final xylene gives a milky haze after coverslipping, not patches without stain.
- C: The hematoxylin was replaced this morning, and weak hematoxylin would not affect the eosin.