Question 1 · Hard
At the morning calibration at 37 °C, an analyzer's ion-selective electrode module rejects the sodium electrode, which shows a slope of 46 mV per tenfold change in activity; the acceptable range is 52 to 68 mV. The potassium slope is 59 mV and the chloride slope -57 mV, both acceptable. For two days, sodium readings had been slow to stabilize and sodium QC had drifted, while potassium and chloride QC stayed near their means. The sodium electrode has been in use for 8 months; the shared reference electrode and the calibrator solutions were replaced last week. The most likely cause is:
- AA protein-coated or aged sodium electrode
- BA blocked junction in the reference electrode
- CContaminated or evaporated calibrator solutions
- DA low temperature in the measuring chamber
Show answerHide answer
Answer: A · A protein-coated or aged sodium electrode
The Nernst equation predicts a slope of about 61.5 mV per tenfold change in activity for a monovalent ion at 37 °C, so 46 mV means the sodium electrode responds poorly. Because only sodium is affected, the fault must lie in the part that sodium does not share with the potassium and chloride electrodes: the sodium-selective membrane. Protein buildup and age reduce a membrane's response, which gives both the slow stabilization and the low slope seen here. The electrode should be cleaned with the manufacturer's deproteinizing solution and recalibrated, and replaced if the slope stays low.
- B: The reference electrode is shared, so a fault there would also affect potassium and chloride.
- C: The calibrators are shared by all the electrodes and would not affect sodium alone.
- D: Temperature changes the slope of every electrode, and only slightly: about 59 mV at 25 °C.