Question 1 · Hard
A small laboratory runs a manual colorimetric phosphorus method for a kidney clinic's batch of samples. The photometer is zeroed on a reagent blank, and five standards give these absorbances: 2.0 mg/dL [0.65 mmol/L], 0.101; 4.0 mg/dL [1.29 mmol/L], 0.204; 6.0 mg/dL [1.94 mmol/L], 0.243; 8.0 mg/dL [2.58 mmol/L], 0.405; 10.0 mg/dL [3.23 mmol/L], 0.508. The package insert claims linearity to 15 mg/dL [4.84 mmol/L], and the controls have not yet been run. The first patient's sample reads 0.300. The technologist should:
- AReport the result read from a line through all points
- BDilute the sample, as the curve bends above 4 mg/dL
- COpen a new reagent lot, as the slope is too shallow
- DRemake the 6.0 mg/dL standard and recalibrate
Show answerHide answer
Answer: D · Remake the 6.0 mg/dL standard and recalibrate
Four of the five standards give about 0.051 absorbance units per mg/dL, a straight line through zero, so the method is linear across the range. The 6.0 mg/dL standard reads 0.243 instead of about 0.305, roughly 20% low, which points to an error in preparing or pipetting that standard rather than a problem with the method. A calibration that contains a clearly wrong point must not be used, and the patient's result sits right where the error matters; read from the four good points, 0.300 would be about 5.9 mg/dL [1.9 mmol/L]. The standard should be remade, the method recalibrated and the controls run before any patient result is reported.
- A: A calibration with a clearly wrong point must not be used; the faulty standard has to be corrected first.
- B: The 8.0 and 10.0 mg/dL standards lie on the line, so the method is linear; only one point is off.
- C: Deteriorated reagent lowers every standard's absorbance, not just one in the middle.