Question 1 · Hard
At 02:40 a technologist verifies a potassium of 6.9 mmol/L on a 71-year-old inpatient with chronic kidney disease; the laboratory's critical limit is 6.0 mmol/L or higher. The hemolysis index is normal, the platelet count is normal, and yesterday's potassium was 5.7 mmol/L, so she releases the result. She phones the ward, where a unit clerk answers, writes 'K 6.9' on a note and promises to pass it on. The technologist records 'critical called to 5 North' in the laboratory system and returns to the bench. At 05:30 the patient has a cardiac arrest, and his nurse says she never received the result. Where did the testing process fail?
- APreanalytical: a new sample was not drawn to confirm it
- BAnalytical: the potassium was not repeated before release
- CPostanalytical: only a physician may receive the result
- DPostanalytical: no caregiver received it with read-back
Show answerHide answer
Answer: D · Postanalytical: no caregiver received it with read-back
The result itself was valid: the sample was not hemolyzed, a normal platelet count makes pseudohyperkalemia unlikely, and a rise from 5.7 mmol/L fits kidney disease. The failure came after analysis, in communication. A critical result must be given directly to a licensed caregiver responsible for the patient, such as the nurse or physician, who reads back the result and the patient identifiers, and the record must show the receiver's name, the time and the read-back. A unit clerk cannot act on the result, and 'called to 5 North' cannot be traced to anyone. If no responsible caregiver can be reached, the laboratory follows its escalation policy, for example calling the physician or the nursing supervisor.
- A: The sample was valid (no hemolysis, normal platelets, a plausible trend), so recollection was not needed.
- B: Repeating a valid critical result is not required and only delays the call; the result was correct.
- C: A responsible licensed caregiver, such as the patient's nurse, may receive and act on a critical result.