Question 1 · Hard
A technologist is subculturing positive blood culture broths in a Class II A2 biological safety cabinet, wearing a gown and gloves. Several open plates and tubes are on the work surface. A tube holding about 10 mL of broth that grew gram-positive cocci in clusters falls over. The broth spreads across the work surface, and some runs through the front grille into the drain pan below. Her gloves are splashed, but her gown and skin are not, and the cabinet's airflow alarm has not sounded. The most appropriate response is to:
- ASwitch the cabinet off and leave the room for 30 minutes
- BClose the sash and run the UV lamp before any cleaning
- CKeep the cabinet running and disinfect it, including the drain pan
- DWipe the surface with alcohol and carry on working
Show answerHide answer
Answer: C · Keep the cabinet running and disinfect it, including the drain pan
A spill inside a working cabinet is already contained: the inflow and downflow keep aerosols away from the operator, so the cabinet is left running. She closes the open plates and tubes, changes gloves, covers the spill with absorbent material soaked in a suitable disinfectant, leaves it for the stated contact time, and wipes the work surface, walls and items inside. Because broth ran through the grille, the drain pan is also flooded with disinfectant, left for the contact time and drained before the surface is wiped again. The cabinet then runs for several minutes before work restarts.
- A: Switching the cabinet off stops the airflow that keeps aerosols inside; leaving the room is for spills outside a cabinet.
- B: UV light does not disinfect liquid spills or shaded areas such as the drain pan.
- D: Alcohol alone is poor on a large protein-rich spill, and the drain pan would stay contaminated.