Question 1 · Hard
An outpatient's blood is collected at 09:00 in a gel separator tube, centrifuged at 09:40 and tested: potassium 4.2 mmol/L (reference 3.5-5.1), with a normal hemolysis index. The capped tube is stored upright at 2-8 °C. Next morning the doctor adds a magnesium request. A small aliquot is first poured off the tube for another add-on test, and its potassium reads 4.3 mmol/L. A processor then re-centrifuges the tube before loading it, and the technologist also repeats potassium on it: 5.5 mmol/L, again with a normal hemolysis index. What best explains the last potassium result?
- ARe-centrifuging the stored gel tube
- BEvaporation from the tube during storage
- CRefrigerating the tube before separation
- DIn vitro hemolysis during overnight storage
Show answerHide answer
Answer: A · Re-centrifuging the stored gel tube
The gel separated the serum from the cells within an hour, and the aliquot poured off next morning still reads 4.3 mmol/L, so overnight storage did not change potassium. The rise appeared only after the tube was spun a second time. Re-centrifuging a stored gel tube lets potassium-rich fluid that has stayed in contact with the cells below the barrier mix with the serum above, which fits a higher potassium with a normal hemolysis index. Gel tubes should not be re-spun: add-on tests use the serum already separated, or a new sample if stability does not allow it.
- B: The tube was capped, and the aliquot taken just before re-spinning shows no rise.
- C: The gel had already separated the serum, and the aliquot taken before re-spinning shows no rise.
- D: The hemolysis index is normal, and the aliquot taken next morning is unchanged.