Question 1 · Hard
A laboratory runs two control levels in every potassium run and rejects any run in which a control falls outside ±2 SD. The limits come from its own 20-day data, and the method's sigma metric is 6.2. In June, 9 of 100 runs were rejected. Each time, the repeated controls fell within 1 SD without any corrective action, the Levey-Jennings charts showed no shift or trend, and EQA results were within 1% of their targets. Patient results were held for about 40 minutes at each rejection. The most likely cause of the rejections is:
- AControl limits that were set too wide
- BFalse rejections expected from a 1-2s rule
- CSystematic error from calibration drift
- DIntermittent random error within the analyzer
Show answerHide answer
Answer: B · False rejections expected from a 1-2s rule
With a Gaussian error distribution, about 4.5% of control results fall outside ±2 SD by chance alone. With two controls per run, 1 - 0.955² ≈ 9% of runs are rejected even when the method is stable, which matches 9 of 100. The clean repeats, flat charts and good EQA results confirm that no real error was present. Using 1-2s as a rejection rule wastes time and teaches staff to ignore QC; for a method near 6 sigma, Westgard sigma rules recommend a single 1-3s rule with two controls, with 1-2s kept, if at all, as a warning.
- A: Limits that are too wide cause fewer rejections, not more.
- C: Calibration drift would show as a shift or trend on the chart and as EQA bias.
- D: Random error would recur on repeat or widen the scatter; every repeat was within 1 SD.