Question 1 · Hard
A 59-year-old woman in the emergency department, breathing 40% oxygen by mask, has an arterial sample taken into a 3-mL plastic syringe with dry balanced heparin. It reaches the laboratory 55 minutes later at room temperature, with a 0.4-mL air bubble trapped under the cap. The doctor wants pH, pCO2, pO2 and lactate to guide treatment for suspected sepsis. The laboratory follows CLSI guidance: air is expelled at once after collection, and plastic syringes kept at room temperature are analyzed within 30 minutes. What should the technologist do?
- AReport the lactate and withhold the gases
- BReject it and request a new anaerobic sample
- CExpel the bubble, remix and analyze it now
- DReport pH and pCO2 and withhold the pO2
Show answerHide answer
Answer: B · Reject it and request a new anaerobic sample
Two separate faults affect every requested result, so the sample cannot be rescued. The trapped air exchanges gas with the blood: CO2 escapes into the bubble, which lowers pCO2 and raises pH, and pO2 moves toward the bubble's value of about 20 kPa (150 mm Hg). During 55 minutes at room temperature the cells keep consuming oxygen and making lactate by glycolysis, so lactate rises and the gases drift further. Expelling the bubble now cannot undo changes that have already happened. The rejection and its reason are documented and the ward is told at once, because a septic patient needs a reliable result quickly.
- A: Lactate keeps rising from glycolysis while blood stands at room temperature, so it is unreliable too.
- C: Gas exchange with the bubble and the long delay have already changed the results; expelling air now cannot reverse them.
- D: The bubble and the delay also change pCO2 and pH, so these results are unreliable as well.