Question 1 · Hard
An AST method follows the fall in NADH absorbance at 340 nm, with malate dehydrogenase as the coupling enzyme, and calculates activity from 60 to 180 seconds after the start reagent is added. Today both QC levels read about 40% below target. The reagent pack has been on board for 6 weeks (stated on-board stability 4 weeks). Raw data for the level 2 control: starting absorbance 1.72 (expected 1.60-1.90); the absorbance then falls by 0.006, 0.011, 0.016 and 0.020 in the four 30-second intervals of the read window. The cuvette temperature log shows 37.0 °C throughout. The most likely cause is:
- AOxidation of NADH in the reagent
- BSubstrate depletion by high AST activity
- CReaction temperature below 37 °C
- DLoss of malate dehydrogenase activity
Show answerHide answer
Answer: D · Loss of malate dehydrogenase activity
In a valid kinetic assay the absorbance falls by the same amount in every interval, because the reaction is zero order and the coupling enzyme turns oxaloacetate into malate as fast as AST makes it. Here the fall grows in each interval, so the reaction is still in its lag phase: too little coupling enzyme is left, oxaloacetate builds up slowly and the measured rate underestimates AST. Malate dehydrogenase loses activity when a reagent is kept past its on-board stability. Oxidized NADH would give a low starting absorbance, and substrate depletion or a cold cuvette would give a slowing or a steady rate, not an accelerating one. Replace the pack, recalibrate and repeat QC before rerunning patients.
- A: Oxidized NADH would lower the starting absorbance, which is normal here.
- B: Substrate depletion makes the rate slow down, not speed up, and needs a very high activity.
- C: The log shows 37.0 °C; a cold cuvette would give a steady but lower rate.