Question 1 · Medium
In capillary zone electrophoresis of serum in an uncoated silica capillary at alkaline pH, the negatively charged proteins still reach the detector near the cathode. Why?
- AThe negative silica wall repels them toward the cathode
- BPressure from the injector pushes them through
- CThe proteins become positive at alkaline pH
- DElectroosmotic flow toward the cathode is stronger
Show answerHide answer
Answer: D · Electroosmotic flow toward the cathode is stronger
At alkaline pH the silanol groups of the silica wall are ionized and negative. Positive buffer ions gather at the wall and are drawn to the cathode, dragging the whole buffer column with them; this is electroosmotic flow. It outweighs the proteins' pull toward the anode, so all proteins are carried to the cathode end, with the most negative ones, such as albumin, arriving last. Capillary electrophoresis gives fast, automated protein separations without staining.
- A: The wall's charge creates electroosmotic flow; it does not push proteins along by repulsion.
- B: Injection pressure loads the sample but does not drive the separation.
- C: Serum proteins are negatively charged at alkaline pH, above their isoelectric points.