Question 1 · Hard
A well outpatient's serum has visible fibrin strands. Indirect ISE on the main analyzer gives sodium 121, potassium 3.4 and chloride 88 mmol/L, and the glucose, urea and total protein from the same run are each about 13% below his results last month. Run from the same tube on the blood gas analyzer, direct ISE gives sodium 139, potassium 3.9 and chloride 101 mmol/L. Measured osmolality is 289 mOsm/kg (reference 275-295) and the serum is clear, not lipemic. QC on both analyzers is within limits. What is the most likely cause of the low results on the main analyzer?
- ATrue hyponatremia from water retention
- BContamination of the sample with IV fluid
- CElectrolyte exclusion by lipid or protein
- DShort sampling by a probe blocked by fibrin
Show answerHide answer
Answer: D · Short sampling by a probe blocked by fibrin
Every result from the main analyzer, electrolytes and photometric tests alike, is low by the same proportion: 121/139, 3.4/3.9 and 88/101 are all about 0.87. That points to less sample than intended being aspirated, which happens when fibrin partly blocks the probe without triggering the clot detector. Electrolyte exclusion needs very high lipid or protein, but the serum is clear and the protein is low, not high. Contamination with IV fluid or true hyponatremia would change the sample itself, so the direct ISE sodium and the osmolality would be low too. Remove the fibrin, re-centrifuge and rerun the sample, and check the probe.
- A: True hyponatremia would give a low direct ISE sodium and a low osmolality.
- B: IV fluid would dilute the whole tube, so direct ISE would also be low.
- C: Exclusion needs high lipid or protein; the serum is clear and the protein is not raised.