Question 1 · Hard
Daily disk diffusion QC with Staphylococcus aureus ATCC 25923 on Mueller-Hinton agar gives these zones, with the acceptable ranges in brackets: gentamicin 16 mm (19-27); erythromycin 19 mm (22-30); tetracycline 33 mm (24-30); cefoxitin 25 mm (23-29); trimethoprim-sulfamethoxazole 27 mm (24-32). The inoculum was a fresh 0.5 McFarland suspension, and the agar depth was 4 mm. A new staff member had incubated the plates for 18 hours at 35 °C in the 5% CO2 incubator used for chocolate agar. What is the most likely cause of the out-of-range results?
- ALow agar surface pH from CO2
- BLoss of potency of the disks
- CExcess thymidine in the agar
- DExcess calcium and magnesium in the agar
Show answerHide answer
Answer: A · Low agar surface pH from CO2
Non-fastidious organisms on Mueller-Hinton agar must be incubated in ambient air, because CO2 dissolves in the agar and lowers its pH. Aminoglycosides and macrolides lose activity at acid pH, so the gentamicin and erythromycin zones shrink, while tetracycline becomes more active and its zone grows. A pattern of some zones too small and one too large points to pH rather than to a general problem such as inoculum or disk potency. Excess divalent cations would shrink the tetracycline zone too, and thymidine would affect trimethoprim-sulfamethoxazole, which is in range. Patient results from this run are held until QC repeated in ambient air is in range.
- B: Weak disks give small zones and cannot explain an oversized tetracycline zone.
- C: Excess thymidine affects trimethoprim-sulfamethoxazole, which is in range here.
- D: Excess cations shrink aminoglycoside and tetracycline zones; they do not enlarge the tetracycline zone.