Question 1 · Medium
A real-time PCR for norovirus is run on a stool specimen, and the amplification curves for the patient and the run controls are shown. The internal control (IC) is valid when its Ct is within 3 cycles of the IC Ct in the negative control. How should the patient result be reported?
- ANorovirus not detected
- BRun failure; repeat all the specimens
- CNorovirus detected, with a high Ct
- DInvalid; re-extract and repeat the test
Show answerHide answer
Answer: D · Invalid; re-extract and repeat the test
The positive control amplifies (Ct about 25) and the negative control shows only its internal control (Ct about 29), so the run is valid. In the patient well the target did not amplify, but the internal control crossed the threshold only at about cycle 37, 8 cycles later than in the negative control, with a low plateau. That delay means something in the stool inhibited the reaction, so a negative target result cannot be trusted. Report the result as invalid, and re-extract or dilute the extract before repeating.
- A: A target-negative result is valid only when the internal control amplifies normally.
- B: Both run controls behaved correctly, so only this specimen is affected.
- C: The late curve is the internal control, not the norovirus target.